Hi @harshk95 Sorry it took me so long to get around to answering this question! These errors show that the start times in the metadata files are not consistent with...
@billkarsh and I actually designed the snsGeomMap partly with SpikeInterface in mind, so that all the information to interpret a SpikeGLX file would available in the metadata, without needing to...
Hi @PathwayinGithub The specific problem you are seeing probably has to do with paths. However, for correct concatenation across multiple runs, you'll need to use the supercat feature in CatGT;...
Two notes about shift register errors: --The error is due to irreversible damage to the shank -- basically some of the electrical contacts to the site have been lost. Once...
It is normal for the NI chassis to not recognize the imec cards -- they never appear in the Ni-MAX listing. They should show up in the Windows Device manager--...
Hi Marc, I really knew from your questions on the Slack channel that you had got past the issue of not seeing the card in Ni-MAX. I don't check the...
Hi Marc, 2.8 V is a little low for TTL (the lowest voltage TTL standard I know is 3.3 V). The SYNC input on the basestation is designed for 5V...
Hi Marc, I see, I just checked the standard threshold for 5V TTL, and see that it is 2.4V. So -- 2.8 should be enough, but since it isn't triggering...
There is no time offset between the sync and neural channels. It is not a bug. It is an effect of the filtering of the AP band. High pass filtering...
Just a couple notes for @Serendipityzzz The NP1110 imro table is completely different than than other NP probes, due to the complicated rules for setting switches. If you need to...