MPR AXIAL image order is reversed

Open liangaili opened this issue 1 year ago • 2 comments

Describe the bug. What happened?

Hi,

I am using weasis 4.4.0 with Windows 11.

Here's what I do:

  1. open a dicom series
  2. open MPR
1

The slice number of a sequence in MPR mode seems reversed: 2

I am not sure if this is a bug or by design.

What version of Weasis are you running?

4.4.0

On which system the problem occurs?

Windows

Relevant log output

No response

Additional contextual elements

No response

liangaili avatar Jul 10 '24 10:07 liangaili

Yes, it is an issue. The sorted order is reversed for all the directions.

nroduit avatar Jul 11 '24 07:07 nroduit

After analysis, this is not a bug but simply a sorting according to Anatomical Direction that is forced in MPR mode.

Anatomical Direction is defined in DICOM the x-axis is increasing to the left hand side of the patient. The y-axis is increasing to the posterior side of the patient. The z-axis is increasing toward the head of the patient.

See also https://github.com/nroduit/Weasis/issues/207

nroduit avatar Aug 15 '24 06:08 nroduit

We're closing this issue due to inactivity over the past several months. If the problem persists or if you have any further information to provide, please feel free to reopen this issue or create a new one. We appreciate your contributions and feedback!