Light sheet mode synchronization issue with Photometrics Iris 15
Cannot optimize remote focus parameters when using the Photometrics Iris 15 in light sheet mode. There are a few problems, which seem related to synchronization:
- Stopping and restarting acquire changes what we see with respect to the remote focusing.
-
- Stop and start acquire:
-
- After setting the parameters where the ASLM almost looks to be working, changing the exposure causes the current remote focus parameters to fail.
-
Camera Configuration
Hi @SJShep, I've made a fix that should solve problem 1. Please test it with the branch #899.
For the waveform parameters, you could try reducing the camera delay time to a shorter value, such as 1 ms. This parameter is set in the configuration.yaml file. At the same time, set an appropriate flyback value and settle-down time for the remote focus through the GUI.
That fixed the problem with stopping and starting the acquisitions.